Notes on X-Headstage IP68 rating

Notes on X-Headstage IP68 rating

The X-Headstage has been tested for IP68 rating by a third party (test report here).  The rating is achieved with the following:
  1. An IP armor is configured with the headstage
  2. The optional moisture resistance coating must be specified at the time of the order. This adds about 100nm of conformal parylene coating to the entire device.
While the headstage can be made and configured to be moisture resistant, care must be taken around both connectors - the system interface connector, the µHDMI or the 12-pin Omnetics, or the probe-facing board-to-board connector.  Silicone can be applied around the connectors once connections are made.  The connection can be disconnected by peeling off the silicone.  If you require more frequent and easier disconnection, X-Headstage connectors can be customized to water-proof type.

If necessary, a thicker (5µm) perylene coating can be applied instead of the standard 100nm. 

Contact us to discuss your needs.

    • Related Articles

    • Notes regarding the reliability of the connectors used on the X-Headstage

      X-Headstage uses ultra-small connectors, called board-to-board connectors, commonly used in the PC industry. These types of connectors have a cycle rating of about 50 cycles according to the manufacturers. But that doesn't mean it will fail at the ...
    • Selecting and Configuring the X-Headstage

      Currently, KONTEX offers several types of X-Headstages X3R32: 32ch recording headstage X3R64: 64ch recording headstage XeR64: e-series 64ch recording headstage XeSR16: e-series 16ch switchable stimulation + recording headstage XeSR32: e-series 32ch ...
    • Notes on X3SR headstage pinout

      X-Headstage is manufactured based on advanced packaging technology such that there are multiple ICs inside each chip. X3SR32 has two Intan RHS2116 dies inside therefore each die is connected to one of the two X3BtB connectors (to maintain the ...
    • Handling the X-Headstage

      Properly handling the X-Headstage will prolong its lifetime and reliability. To plugin connect the EIB/Probe/Adapter, align the two units directly on top of each other and exert equal force along the entire area. If the connectors are plugged in ...
    • Is armor necessary for the X-Headstage?

      It is possible to use the X-Headstage without the armor, but having an additional layer will offer additional mechanical protection and will likely prolong its lifetime.