XDAQ HDMI Port and Supported X-Headstage Explained
The XDAQ line supports both Record (X3R & X6R) and Stim-Record (X3SR) X-Headstages (as well as other Intan-compatible headstages). Depending on the purchased configurations, some headstage HDMI ports in the XDAQ might not be enabled by default. The rule of thumb is to use Port 1 and other ports if necessary.
Here is a breakdown of the number of channels and ports supported by the XDAQ CORE and XDAQ ONE models.
XDAQ CORE
SKU: XDAQ-CORE_S16 - supports total of 512 Rec channels and 16 Stim-Rec channels
SKU: XDAQ-CORE_S32 - supports total of 512 Rec channels and 32 Stim-Rec channels
| XR Mode | XSR Mode |
|
| -R512 | -S16 | -S32 |
Port 1 | 128ch or 2 RHD chips | 16ch or 1 RHS chip | 32ch or 2 RHS Chips |
Port 2 | 128ch or 2 RHD chips | non-functional | non-functional |
Port 3 | 128ch or 2 RHD chips | non-functional | non-functional |
Port 4 | 128ch or 2 RHD chips | non-functional | non-functional |
XDAQ ONE
SKU: XDAQ-ONE_R512_S64 - supports total of 512 Rec channels and 64 Stim-Rec channels
SKU: XDAQ-ONE_R512_S128 - supports total of 512 Rec channels and 128 Stim-Rec channels
SKU: XDAQ-ONE_R1024_S64 - supports total of 1024 Rec channels and 64 Stim-Rec channels
SKU: XDAQ-ONE_R1024_128 - supports total of 1024 Rec channels and 128 Stim-Rec channels
| XR Mode |
| XSR Mode |
|
| -R512 | -R1024 | -S64 | -S128 |
Port 1 | 256ch or 4 RHD chips | 256ch or 4 RHD chips | 32ch or 2 RHS Chips | 32ch or 2 RHS Chips |
Port 2 | 256ch or 4 RHD chips | 256ch or 4 RHD chips | 32ch or 2 RHS Chips | 32ch or 2 RHS Chips |
Port 3 | non-functional | 256ch or 4 RHD chips | non-functional | 32ch or 2 RHS Chips |
Port 4 | non-functional | 256ch or 4 RHD chips | non-functional | 32ch or 2 RHS Chips |
Note that it is not possible, for example, to upgrade XDAQ CORE to drive 64ch of Stim-Rec X-Headstages. Finally, it is not possible to drive both R and SR X-Heastages at the same time.
Related Articles
XDAQ supported headstages
In addition to the ultra-small X-Headstages, XDAQ supports existing Intan-compatible headstages (from Intan, NeuroNexus, and more). To run these headstages with their corresponding accessories, you will need an adapter to properly remap the control ...
How to run more than 4 headstages with the XDAQ?
XDAQ is designed for high-channel neural acquisition. Although each box has only 4 ports on board, it is possible to drive more than 4 headstages/animals at the same time (though each with a lower channel count configuration). For example, each XDAQ ...
What is the difference between XDAQ Core and One?
The XDAQ Core is the entry-level box supporting electrical stimulation and high-speed neural recording. It can drive up to 512ch of recording X-Headstages or up to 32ch of stim-record X-Headstages. XDAQ One is the solution for users needing higher ...
XDAQ DB25 pinout
DB25 connector pin out for XDAQ ONE and CORE DB Pin Signal 1 2 DI 03 3 DI 05 4 DI 07 5 6 7 8 9 10 DO 04 11 DO 06 12 DO 08 13 DAQ ON 14 SYS GND 15 DI 04 16 DI 06 17 DI 08 18 19 20 21 22 DO 03 23 DO 05 24 DO 07 25 Sample CLK DAQ ON: logic high at onset ...
What do the XDAQ Status LEDs mean?
XDAQ CORE LED Indicators SYS [ON] Program Loaded [Flashing] Hardware Error S1 [ON] Acquisition ON S2 [Flashing] Activities* in DIN1 S3 [Flashing] Activities* in DOUT1 P1 [ON] Headstage Detected on Port 1 (after software initialization) P2 [ON] ...