Most high (1-5MOhm) or low (1-100kOhm) impedance electrodes, commercially available or homemade, can be used with the X-Headstage. Be mindful that for stimulation experiments, high impedance electrodes might not be suitable as x3SR has a compliance stim voltage of 14V. It is recommended to condition the electrode by lowing its impedance through electroplating, activation or other techniques before proceeding with stimulation experiments.